Dear Customers, We would like to inform you that The
Transkrypt
Dear Customers, We would like to inform you that The
Dear Customers, We would like to inform you that The Research and Development Department of PZ CORMAY S.A. introduced the modification of the application for ALP II GENERATION for Prestige 24i and Biolis 24i Premium. The change concerns the new absorbance limits: Abs. Limit Previous version Current version Low -0.1 -3 High + 0.8 +3 This modification allows to eliminate appearing flag "G" with test results. The modification will not affect the results of alkaline phosphatase in the sample. Please find enclosed the actual instruction for the ALP II GENERATION. Please read the reagent manual carefully and correct absorbance limits range in software of the analyzer (test item parameter screen). Best regards, Marzena Gałczyńska Product Manager ds. biochemii W razie pytań jestem do dyspozycji: tel.: 81 749 44 60 e-mail: [email protected] tel. kom.: 661 551 196 WI 1510/01